Instrument type: Fully automatic, discrete, random selection, emergency priority biochemical analyzer Colorimetric method: Direct measurement in reaction cup Analysis method: 1. End point method, kinetic method, two-point method , dual-wavelength method, immunoturbidimetric method, dual-reagent method, serum blank method, multi-standard and other methods;
2. All detection methods are open, and imported or domestic reagents can be used. Analysis items: Online biochemical testing up to 40 items Maximum speed: Comprehensive testing speed 1200 tests/hour hba1c test: 400 tests/hour (including hemolysis step) Sample device: 1. 60 sample positions, including standard, quality control, and emergency positions ;
2. 2 virtual sample trays can be set up, and 180 sample working units can be arranged at one time;
2. Original test tubes (vacuum negative pressure blood collection tubes) or serum can be used Cup with sample refrigeration function. Sample volume: 1-100μL, equipped with high-precision diluter. Reagent positions: 1. 40 refrigerated reagent positions, which can be expanded to 80 reagent positions virtually
2. The instrument is equipped with a reagent refrigeration module as standard
3. Split design is also available Overall refrigerator refrigerated reagent volume: R1: 1-400μL; R2: 1-400μL Liquid level detection: Yes, it can automatically detect the remaining amount of reagents in the reagent bottles to ensure minimal contact between reagents and samples and reduce cross-contamination; it has a volume tracking function , the sampling needle has an automatic maintenance and cleaning function. Response time: 0-900 seconds, and the time can be set. Colorimetric cup: 90 miniature high-quality transparent purple plexiglass reaction cups, reusable, low consumption, and low cost.
(Hard quartz glass is also optional) Minimum reaction volume: 180μL Reaction temperature: 37℃±0.1℃ Cleaning: 1. Automatic 8-stage flushing, special pollution items can choose 8-stage 16-time cleaning , effectively solve cross-contamination;
2. Waste liquid can be collected separately in high concentration and low concentration; water consumption is about 3 liters per hour;
3. The cuvette is automatically dried. Alarm function: Insufficient reagents and samples; lack of water; overflow of waste liquid; Calibration: Linear/nonlinear multi-point calibration sample dilution/retest: When the result exceeds the linear range, the instrument can automatically dilute the sample and retest; also Test after dilution can be preset; and has automatic cycle dilution function Wavelength range: 1. 340-800nm ??(standard configuration is 340-700nm, can be increased or decreased according to customer needs)
2. Adopt post-spectroscopy Technology, long-life halogen lamp detection range: 0-4.00.0D Instrument accuracy: accurate to 0.0001O.D, multiple calculation methods are set, and users can set the system calibration according to their needs: blank bottom value deduction, automatic zero adjustment, instrument The blank value of each cuvette can be automatically recorded and deducted to obtain the true result, ensuring reliable and accurate results. The instrument also has an automatic light drift elimination function to ensure long-term stable operation of the instrument. Emergency function: Emergency samples can be inserted at any time, and standard/quality control and reagent blanks can also be inserted at any time. Stirring function: It has an independent stirring needle to ensure uniform mixing and accurate results. Safety guarantee: This product is carefully designed and manufactured in accordance with international standards and uses a number of patented technologies to achieve high detection accuracy. The instrument is equipped with functions such as fault alarm, automatic diagnosis and quality control to ensure that the test results are accurate and reliable. Quality Control: It has rich quality control processing functions. Sample quality control standards can be added to the test at will without stopping the machine; the entire process displays the reaction curve (including the sample quality control standard curve) in real time, and can recalculate the analysis; the quality control chart can be stored, displayed, and printed. Different quality control substances can be predefined, each test can have multiple quality control substances at the same time, and multiple modes of quality control charts can be selected. Software: 1. Runs under Windows , line-log and other response curves, and has a variety of calculation methods. You can set a variety of calculation formulas by yourself and set and select other international languages.
Data processing: 1. More than 900 detection methods can be edited and stored for a long time;
2. The instrument can store more than 1 million patient data for a long time. The curve of the entire reaction process is saved and can be reanalyzed and calculated without retesting. Printing: Multiple printing formats, you can choose Chinese (English) report form; users can modify the printing format by themselves Power supply: AC220V 50Hz 1000VA Volume: 88*73*109cm