Introduction to vortex mixer

The multi-module vortex mixer is small in size and can mix quickly and thoroughly. It can mix all the test solutions attached to the tube wall. It is suitable for general test tubes, centrifuge tubes, liquid separation, and liquid separation in funnels. Mixes evenly and is widely used in biochemistry, genetic engineering, medicine and other experimental needs. Lichen Technology focuses on the field of scientific services, providing one-stop laboratory products and services, including general instruments, analytical instruments, physical property testing, environmental testing, instrument consumables, etc.

1. Product Features

1. The large LCD screen displays the speed and time, which is convenient for setting and reading; 2. Suitable for short-term (jogging) or long-term continuous work; 3. , DC brushless motor, long life, strong power, maintenance-free; 4. Special vibration head, strong toughness, sufficient elasticity, long service life; 5. Wide speed range, maximum speed 3000rpm, stepless speed regulation; 6. Module installation Convenient, the eccentric bearing design is durable; 7. Using high-quality foot pads, the instrument has good stability and low operating noise.

2. Equipment installation

1. Selection of working environment? (Attention! Please select the working environment according to the requirements below)

(1) The working environment should Keep it clean and dry; (2) Avoid direct sunlight, severe temperature fluctuations and air convection; (3) Keep it as far away from doors, windows, radiators and air outlets of air conditioning units as possible; (4) The instrument should be placed on a level, stable surface on the workbench; (5) the workbench should be set up in a place that is less susceptible to vibration interference; (6) the instrument should be kept away from objects and equipment that are magnetic or can generate magnetic fields; (7) not in areas with explosive hazards Use the instrument; (8) Do not use the instrument in an environment with high humidity or high dust for a long time.

2. Equipment installation

(1) After unpacking, remove all packaging and remove the buffer sponge; (2) Place the mixer on a stable, vibration-free workbench. As level as possible; (3) Plug the output end of the power adapter into the power socket at the rear of the instrument

3. Operation method:

Ordinary model

Inching Mode:

1. Press the switch to the jog mode, the jog indicator light on the panel lights up, and the speed cannot be adjusted; 2. Place the mixed sample tube in the center of the standard head, and allow the sample tube to Keep it vertical; 3. Gently press the sample tube to make the standard head start to oscillate; 4. After mixing is completed, separate the sample tube from the standard head and the instrument stops running.

Continuous operation mode:

1. Replace the standard head with a universal clamp and tighten the screws; 2. Insert the centrifuge tube containing the mixed sample into the test tube evenly module, the mixed samples in each test tube should be roughly equal, put the test tube module into the universal clamp and fasten it; 3. Press the switch to the continuous operation mode, the continuous operation light will light up, and the speed cannot be adjusted; 4. The instrument starts running and the operation is completed. Finally, the instrument turns off when the switch is in the middle position.

Speed ??adjustment model:

Inching mode:

1. Click the "Touch/Cont" button to switch to inching mode, and the inching indication on the panel The light lights up and the speed is adjustable; 2. Turn the speed adjustment knob counterclockwise to the lowest speed; 3. Place the mixed sample tube in the center of the standard head and keep the sample tube vertical; 4. Gently press the sample tube while Slowly rotate the speed knob to the required speed, and the standard head begins to oscillate; 5. After mixing is completed, separate the sample tube from the standard head, and the instrument stops running.

Continuous operation mode:

1. Replace the standard head with a universal clamp and tighten the screws; 2. Insert the centrifuge tube containing the mixed sample into the test tube evenly Clamp, the mixed sample in each test tube should be roughly equal, put the test tube clamp into the universal clamp and fasten it; 3. Turn the speed adjustment knob counterclockwise to the lowest speed; 4. Click the "Touch/Cont" button to switch to continuous operation mode , the continuous running light lights up, and the speed is adjustable; 5. Slowly rotate the speed control knob, set to the required speed, and the instrument starts running; 6. After the operation is completed, turn off the power of the instrument when the switch is in the middle gear.

Digital display model:

Jog mode:

1. Click the "Touch/Cont" button to switch to jog mode, and the jog indication on the panel The light lights up; 2. Press the knob to set the parameters, rotate the knob, and set to the required speed; 3. Place the sample tube with mixing in the center of the standard head, and keep the sample tube vertical; 4. Press the sample tube gently, The standard head starts to oscillate, the rotation speed rises to the set speed, and the cumulative timing begins; 5. After mixing is completed, separate the sample tube from the standard head, the instrument stops running, and the timing is cleared.

Continuous operation mode:

1. Replace the standard head with a universal clamp and tighten the screws; 2. Insert the centrifuge tube containing the mixed sample into the test tube evenly Clamp, the mixed sample in each test tube should be roughly equal, put the test tube clamp into the universal clamp and fasten it; 3. Turn the speed adjustment knob counterclockwise to the lowest speed; 4. Click the "Touch/Cont" button to switch to continuous operation mode , the continuous running light lights up, and the speed is adjustable; 5. Rotate the speed control knob and set it to the required speed; 6. Click the speed control knob to switch to timing mode, and rotate the knob to set the required timing duration; 7. Click "Stop/ Run" key, the instrument starts running, the speed rises to the set speed, and the countdown begins. If the timing is "0", it will run for a long time; 8. After the operation is completed, the instrument will automatically stop running. If the timing is "0", you need to click the "Stop/Run" button to stop the instrument.

IV. Maintenance

1. Keep the instrument clean to prevent the solution from flowing into the instrument; 2. Wipe the vibration head, body and feet with a soft cloth dipped in neutral detergent. pad, and then use a soft cloth filled with distilled water to remove the detergent; 3. When not using it for a long time, please power off the instrument and place it together with the instrument accessories in a dust cover for safekeeping.