Parameter list of automatic biochemical analyzer

Instrument type: automatic, discrete, random and emergency priority biochemical analyzer Colorimetry: reaction cup direct determination and analysis methods: 1, endpoint method, kinetic method, two-point method, dual-wavelength method, immunoturbidimetry, double reagent method, serum blank method, multi-standard method, etc.

2. All detection methods are open, and imported or domestic reagents are acceptable. Analysis items: online biochemical detection can reach up to 40 speeds: constant speed 300 times/hour (or ISE500 times/hour) Sample device: 1, 60 sample positions, including standard, quality control and emergency position;

2. Two virtual sample trays can be set, and 180 sample work units can be placed at one time;

2. The original test tube (vacuum negative pressure blood collection tube) or serum cup can be used, which has the function of sample refrigeration. Sample size: 1- 100μL, equipped with high-precision diluter. Reagent sites: 1, 40 cold storage reagent sites, which can be virtually expanded to 80 reagent sites.

2. The instrument comes standard with reagent refrigeration module.

3. Split design can also be used in the whole refrigerator. Test dose: r1:1-400μ l; R2: 1-400μL liquid level detection: it can automatically detect the remaining reagents in reagent bottle, ensure the minimum contact between reagents and samples, and reduce cross-contamination; With follow-up function, the sampling needle has automatic maintenance and cleaning function. The reaction time is 0-900 seconds, and the time can be set: 90 miniature high-quality purple transparent plexiglass reaction cups can be reused, with low consumption and cost.

(Hard seasonal glass can also be selected) Minimum reaction amount: 180μL reaction temperature: 37℃ 0. 1℃ cleaning:1,automatic 8-stage washing, and 8-stage 16 cleaning can be selected for special pollution projects, effectively solving cross-contamination;

2. The waste liquid can be collected with high concentration and low concentration respectively; The water consumption is about 3 liters per hour;

3. The cuvette will automatically dry. Alarm function: insufficient reagents and samples; Lack of water; Waste liquid overflows; Calibration: linear/nonlinear multi-point calibration sample dilution/re-inspection: when the result exceeds the linear range, the instrument can automatically dilute and re-inspect the sample; The diluted test can also be preset; And it has the function of automatic circulation dilution: wavelength range: 1, 340-800nm (340-700 nm is the standard, which can be increased or decreased according to customer requirements).

2. Using post-splitting technology, the detection range of long-life halogen lamp is 0-4.00.0D, and the instrument accuracy is 0.0001o.. D At the same time, users can set the system calibration by themselves according to their needs: blank bottom value deduction, automatic zero adjustment, and the instrument can automatically record and deduct the blank value of each cuvette, so as to obtain real results and ensure the reliability and accuracy of the results. The instrument also has the function of automatically eliminating light drift to ensure the long-term stable operation of the instrument. Emergency function: emergency samples can be inserted at any time, and standard/quality control and reagent blanks can also be inserted at any time. Mixing function: it has an independent mixing needle to ensure even mixing results and accurate safety guarantee: this product is carefully designed and manufactured according to international standards, and adopts a number of patented technologies with high detection accuracy. The instrument has the functions of fault alarm, automatic diagnosis and quality control to ensure the accuracy and reliability of the test results. Quality control: It has rich quality control and processing functions. Sample quality control standards can be added to the test at will without stopping the machine; The reaction curve (including the sample quality control standard curve) can be displayed in real time, and the analysis can be recalculated; The quality control chart can be stored, displayed and printed. Different quality controls can be pre-defined, and multiple quality controls can be brought at the same time for each test, and multiple quality control charts can be selected. Software: 1, running in Windows XP environment, with Chinese and English versions available;

2. Rich query function; The interface is rich in display and easy to use;

3. It has line-line, line-logarithm and other response curves. , and there are a variety of calculation methods, you can set up various calculation formulas and choose other international languages. Data processing: 1, more than 900 detection methods can be edited and saved for a long time;

2. The instrument can store more than 6.5438+0 million patient data for a long time. Save the curve of the whole reaction process, and re-analyze and calculate without re-measurement. Print: multiple print formats, Chinese (English) reports can be selected at will; Users can modify the print format by themselves. Power supply: AC220V 50Hz 1000VA Volume: 88*73* 109cm Registration number: Guo Yao Guan Xie (Jin) Zi 2002 No.2400279 Manufacturer's name (Chinese) Product performance structure and composition: portable, semi-automatic analyzer, controller and analyzer integrated; Analysis methods: endpoint method, two-point method and dynamic method; Wavelength range: 340-750 nm. The center wavelengths of the interference filter are marked as 340, 405, 505, 600 and 750 nanometers. Scope of application This biochemical analyzer is used for the analysis of clinical biochemical items. The registered agent is Beijing Huashixin Medical Equipment Co., Ltd., the after-sales service organization of Beijing Huashixin Medical Equipment Co., Ltd. Approval date 2002.04.0 1 Expiry date 2006.04.065438 Manufacturer's name (English) UV-VIS Metrolab S.A Manufacturer's address (Chinese) Misions150.