What does the power supply e 1 of Bole Electrophoresis Instrument mean?

It shows that the power supply failed to form a loop with the electrophoresis tank after being electrified. That is, there is an open circuit between the electrophoresis tank and the power supply.

There are two reasons:

1, the connection line between the power supply and the electrophoresis tank is not well connected, that is, the power supply is not energized;

2. The buffer solution in the electrophoresis tank leaked to the outside of the electrophoresis module, that is, the buffer solution in the upper pool could not contact with the electrophoresis gel, resulting in the failure to form a power-on loop.

Checking and solving the above two reasons can eliminate the problem. LAB-EYE, a creative creature, hopes to help you.