Shimadzu GC-17A technical description
1. GC-17A Ver.3 can provide complete digital flow path control, including: diaphragm purge flow path, etc. GC-17A is The first instrument in the world to provide digital setting of split ratio. (U.S. Patent: US5339673, Japanese Patent: JP2131776)
2. GC-17A Ver.3 can install 3 inlets and 4 detectors at the same time. Two automatic samplers can be installed at the same time.
3. GC-17A Ver.3 PTV and OCI are the same inlet, and the conversion from PTV to OCI can be completed using the keyboard.
4. GC-17A Ver.3 can set a week's program for automatic analysis.
5. GC-17A Ver.3 has the function of counting the number of injections, which can determine the replacement cycle of the injection pad. In addition, it also has the function of automatic statistics of coolant consumption. These all support GLP/GMP hardware requirements.
6. GC-17A Ver.3 has a SYSTEM key, which can be used to turn off all heating power. In addition, you can use the time program to automatically turn off the system keys to achieve