1, fully preheat before use, generally more than 30 minutes. The higher the precision of the balance, the longer the preheating time should be.
2. Calibrate with standard weights before use, and select weights of corresponding grades for scales with different accuracy.
3. Readings should be taken at a fixed time after loading. For example, all readings should be taken at the indicated value of 5 seconds after loading or when all readings are displayed in units.
4, multiple measurements, average (3~5 times).
5, the room should avoid direct sunlight, it is best to choose a shady room or use shading.
6. The precision balance should be far away from sound sources, such as railways, highways, vibrating machines and other vibrating machinery. When it is unavoidable, shockproof measures should be taken.
7. Stay away from heat sources and high-intensity electromagnetic fields.
8. The temperature of the studio should be constant, preferably around 20℃.
9. The relative humidity of the studio should be between 45% and 75%.
10, the studio should be clean to avoid the influence of airflow.
1 1. Before use, adjust the level to the middle position.